A Mueller bridge set for cryogenic temperature measurements

From

Risø National Laboratory, Risø National Laboratory for Sustainable Energy, Technical University of Denmark1

Risø National Laboratory for Sustainable Energy, Technical University of Denmark2

An a.c. Mueller bridge set for resistance thermometry at cryogenic temperature is described. A commercial tuned null detector is used at an operating frequency of 1025 c/s. The set includes a high stability oscillator, line reject filter, phase shifter, Q multiplier and selector box. The latter permits the dissipation in the thermometers not being measured to be maintained at the operating level with direct current.

A temperature change of the order of 10 μdegK can be detected with 10-8 W applied to the thermometer.

Language: English
Year: 1966
Pages: 576-580
ISSN: 1747387x and 09507671
Types: Journal article
DOI: 10.1088/0950-7671/43/8/329