About

Log in?

DTU users get better search results including licensed content and discounts on order fees.

Anyone can log in and get personalized features such as favorites, tags and feeds.

Log in as DTU user Log in as non-DTU user No thanks

DTU Findit

Journal article

Some Limitations in the Use of Plastic and Dyed Plastic Dosimeters

From

Risø National Laboratory for Sustainable Energy, Technical University of Denmark1

Several practical plastic and dyed plastic dosimeters were examined under irradiation conditions similar to those used for radiation processing of materials. Cellulose triacetate, polymethyl methacrylate, polyvinyl chloride, dyed polymethyl methacrylate, dyed Cellophane and dyed Nylon were given fractionated and uninterrupted absorbed doses in the megarad range with 60Co γ-rays and 10 MeV electron beams.

It was found that with some systems, differences in radiation response due to dose-rate and temperature dependence can cause large systematic errors in dose interpretation. Poor reproducibility of response may result from batch-to-batch differences or intra-batch variations in thickness, distribution of sensitizers and background optical density.

When using these routine dosimeters, careful calibration of the response and monitoring of factors contributing to poor reproducibility are essential.

Language: English
Year: 1975
Pages: 611-620
ISSN: 18781284 and 0020708x
Types: Journal article
DOI: 10.1016/0020-708X(75)90079-4
ORCIDs: Miller, Arne

DTU users get better search results including licensed content and discounts on order fees.

Log in as DTU user

Access