About

Log in?

DTU users get better search results including licensed content and discounts on order fees.

Anyone can log in and get personalized features such as favorites, tags and feeds.

Log in as DTU user Log in as non-DTU user No thanks

DTU Findit

Journal article

Highly selective micro-sequential injection lab-on-valve (μSI-LOV) method for determination of ultra trace concentrations of nickel in saline matrices using detection by electrothermal atomic absorption spectrometr

From

Department of Chemistry, Technical University of Denmark1

University of the Balearic Islands2

Technical University of Denmark3

A highly selective procedure is proposed for the determination of ultra trace level concentrations of nickel in saline aqueous matrices exploiting a micro sequential injection lab-on-salve (μSI-LOV) sample pretreatment protocol comprising bead injection separation/pre-concentration and detection by electrothermal atomic absorption spectrometry (ETAAS).

Based on the dimethylglyoxime (DMG) gravimetric procedure used for nickel analysis, the sample, as contained in pH 9.0 buffer, is, after on-line merging with the chelating reagent, transported to a reaction coil attached to one of the external ports of the LOV to assure sufficient reaction time for the formation of Ni(DMG)2 chelate.

The non-ionic coordination compound is then collected in a renewable micro-column packed with a reversed-phase copolymeric sorbent [namely, poly(divinylbenzene-co-N-vinylpyrrolidone)] containing a balanced ratio of hydrophilic and lipophilic monomers. Following elution by a 50 μL methanol plug in an air-segmented modality, the nickel is finally quantified by ETAAS.

Under the optimized conditions and for a sample volume of 1.8 mL, a retention efficiency of 70 % and an enrichment factor of 25 were obtained. The proposed methodology showed high tolerance to the commonly encountered alkaline earth matrix elements in environmental waters, that is, calcium and magnesium, and was successfully applied for the determination of nickel in an NIST standard reference material (NIST 1640-Trace elements in natural water), household tap water of high hardness and local seawater.

Satisfying recoveries were achieved for all spiked environmental water samples with maximum deviations of 6%. The experimental results for the standard reference material were not statistically different to the certified value at a significance level of 0.05.

Language: English
Publisher: Springer-Verlag
Year: 2006
Pages: 739-748
ISSN: 16182650 and 16182642
Types: Journal article
DOI: 10.1007/s00216-006-0467-5