The effect of irradiation temperatures between ambient and 80 deg. C on the response of alanine dosimeters

From

National Physical Laboratory1

Radiation Physics, Radiation Research Division, Risø National Laboratory for Sustainable Energy, Technical University of Denmark2

Radiation Research Division, Risø National Laboratory for Sustainable Energy, Technical University of Denmark3

Risø National Laboratory for Sustainable Energy, Technical University of Denmark4

Published data on the effect of irradiation temperature on the response of alanine dosimeters does not extend to the temperatures that may be experienced in high-dose industrial irradiations, particularly in the case of electron beams. We describe here results of the irradiation of alanine dosimeters at temperatures up to 80 °C and doses up to 70 kGy.

Data have been obtained for both 60Co and electron beam irradiations and the effect of temperature on the stability of the radiation-induced signal has also been investigated. At temperatures above 50 °C the irradiation temperature coefficient begins to deviate significantly from linearity and shows marked dose dependence.

The effect of this behaviour under conditions typically experienced in industrial processing is evaluated and recommendations made concerning the use of alanine dosimeters at high doses and temperatures.

Language: English
Year: 2009
Pages: 473-475
ISSN: 18790895 and 0969806x
Types: Journal article
DOI: 10.1016/j.radphyschem.2009.03.028