Journal article

Diagnostic PCR: Comparative sensitivity of four probe chemistries

From

Division of Microbiology and Risk Assessment, National Food Institute, Technical University of Denmark1

National Food Institute, Technical University of Denmark2

Three probe chemistries: locked nucleic acid (LNA), minor groove binder (MGB) and Scorpion were compared with a TaqMan probe in a validated real-time PCR assay for detection of food-borne thermotolerant Campylobacter. The LNA probe produced significantly lower Ct-values and a higher proportion of positive PCR responses analyzing less than 150 DNA copies than the TaqMan probe.

Choice of probe chemistry clearly has an impact on the sensitivity of PCR assays, and should be considered in an optimization strategy.

Language: English
Year: 2009
Pages: 201-203
ISSN: 10961194 and 08908508
Types: Journal article
DOI: 10.1016/j.mcp.2009.02.003
ORCIDs: Löfström, Charlotta and Hoorfar, Jeffrey