About

Log in?

DTU users get better search results including licensed content and discounts on order fees.

Anyone can log in and get personalized features such as favorites, tags and feeds.

Log in as DTU user Log in as non-DTU user No thanks

DTU Findit

Journal article

An enzymatic nucleic acid vertical flow assay

From

Tarbiat Modarres University1

Department of Biotechnology and Biomedicine, Technical University of Denmark2

Section for Microbial and Chemical Ecology, Department of Biotechnology and Biomedicine, Technical University of Denmark3

Nano Bio Integrated Systems, Section for Microbial and Chemical Ecology, Department of Biotechnology and Biomedicine, Technical University of Denmark4

Vertical flow assays have been developed in recent years addressing limitations of the lateral flow assays, including limited multiplexing capability, quantitation, and hook effect. In the present study, the first passive paper-based vertical flow assay is developed for the detection of the nucleic acid target.

Horseradish peroxidase was used as an enzymatic tracer with a high potential for signal amplification. In order to achieve the best signal-to-noise ratio, different parameters of paper-based assays were optimized. The sample is heat denatured and hybridized with a specific probe to form a dual-labeled hybridization complex.

A small volume of diluted sample, 12 µl, can be analyzed within 6 min on the assay in a sandwich format. Assay specificity was evaluated by testing different unrelated samples, and also, 1.7 nM was obtained as the limit of detection (LOD) using the 0 + 3SD method, which is equivalent to 8.5 fmol of double-stranded DNA in the 12 µl sample volume.

The linear range of 3–194 nM with a 0.978 correlation coefficient was obtained according to the calibration curve. The developed assay was evaluated with 45 hepatitis B virus clinical plasma samples, and the result showed 100% consistency of the assay with the real-time PCR benchmark. In the present study, we sought to develop a mere detection system for nucleic acid targets, and to investigate the possibility of using enzyme reporter in a passive vertical flow assay.

Graphical abstract: [Figure not available: see fulltext.]

Language: English
Publisher: Springer Berlin Heidelberg
Year: 2022
Pages: 3605-3615
ISSN: 16182650 and 16182642
Types: Journal article
DOI: 10.1007/s00216-022-03988-7
ORCIDs: Svendsen, Winnie Edith

DTU users get better search results including licensed content and discounts on order fees.

Log in as DTU user

Access

Analysis