Journal article

Filament stretching rheometer: inertia compensation revisited

From

Department of Chemical and Biochemical Engineering, Technical University of Denmark1

Massachusetts Institute of Technology2

The necessary inertia compensation used in the force balance for the filament stretching rheometer is derived for an arbitrary frame of reference. This enables the force balance to be used to extract correctly the extensional viscosity from measurements of the tensile force at either end of the elongating fluid column in any of the different experimental configurations that have been Introduced to date.

The present analysis eliminates a restriction inherent in the work of Szabo.

Language: English
Publisher: Springer-Verlag
Year: 2003
Pages: 269-272
ISSN: 14351528 and 00354511
Types: Journal article
DOI: 10.1007/s00397-002-0277-4