Conference paper
Evaluation of a novel PTFE material for separation and preconcentration of trace levels of metal ions in sequential injection (SI) and sequential injection lab-on-valve (SI-LOV) systems interfaced with detection by ETAAS
The operational characteristics of a novel PTFE bead material, granular Algoflon®, used for separation and preconcentration of metal ions via adsorption of on-line generated non-charged metal complexes, were evaluated in a sequential injection (SI) system furnished with an external packed column and in a sequential injection lab-on-valve (SI-LOV) system.
Employed for the determination of cadmium(II), complexed with diethyldithiophosphate (DDPA), and detection by electrothermal atomic absorption spectrometry (ETAAS), its performance was compared to that of a previously used material, Aldrich PTFE, which had demonstrated that PTFE was the most promising for solid-state pretreatments.
By comparing the two materials, the Algoflon® beads exhibited much higher sensitivity (1.6107 versus 0.2956 μg l-1 per integrated absorbance (s)), and better retention efficiency (82% versus 74%) and enrichment factor (20.8 versus 17.2), although a slightly smaller linear dynamic range (0.05-0.25 μg l-1 versus 0.05-1.00 μg l-1).
Moreover, no flow resistance was encountered under the experimental conditions used. The results obtained on three standard reference materials were in good agreement with the certified values.
Language: | English |
---|---|
Year: | 2004 |
Proceedings: | 8. Danske Symposium i Analytisk Kemi |
Types: | Conference paper |