Range/velocity limitations for time-domain blood velocity estimation

From

Department of Electrical Engineering, Technical University of Denmark1

Biomedical Engineering, Department of Electrical Engineering, Technical University of Denmark2

The traditional range/velocity limitation for blood velocity estimation systems using ultrasound is elucidated. It is stated that the equation is a property of the estimator used, not the actual physical measurement situation, as higher velocities can be estimated by the time domain cross-correlation approach.

It is demonstrated that the time domain technique under certain measurement conditions will yield unsatisfactory results, when trying to estimate high velocities. Various methods to avoid these artifacts using temporal and spatial clustering techniques are suggested. The improvement in probability of correct detection is derived, and several examples of simulations are shown.

Language: English
Publisher: Elsevier BV
Year: 1993
Pages: 741-749
ISSN: 1879291x and 03015629
Types: Journal article
DOI: 10.1016/0301-5629(93)90091-2
ORCIDs: Jensen, Jørgen Arendt