Journal article
Determination of urea, glucose, and phosphate in dialysate with Fourier transform infrared spectroscopy
Individual control and quantification of phosphate removal is desirable in dialysis treatment. Currently, no on-line method exists to quantify phosphate removal. We demonstrate that a multivariate calibration model based on infrared transmission spectra is capable of predicting phosphate, urea, and glucose concentrations at clinically relevant levels.
The on-line monitoring of these components by infrared spectroscopy is therefore feasible.
Language: | English |
---|---|
Year: | 2004 |
Pages: | 899-905 |
ISSN: | 18733557 and 13861425 |
Types: | Journal article |
DOI: | 10.1016/S1386-1425(03)00317-2 |