Conference paper
Investigating a CMUT's Ability to Achieve Non-linear Contrast Enhancement
Center for Fast Ultrasound Imaging, Biomedical Engineering, Department of Health Technology, Technical University of Denmark1
Biomedical Engineering, Department of Health Technology, Technical University of Denmark2
Department of Health Technology, Technical University of Denmark3
Mems Applied Sensors Group, Biomedical Engineering, Department of Health Technology, Technical University of Denmark4
Capacitive micromachined ultrasonic transducers (CMUTs) are inherently non-linear and often believed to be unsuitable for non-linear contrast-enhanced ultrasound (CEUS) imaging. The ability of CMUTs to perform contrast-enhanced ultrasound imaging has been investigated by experimentally evaluating the contrast-to-noise ratio (CTR) and its variation with emitted pressure.
This has been evaluated using amplitude modulation, pulse inversion and B-mode imaging sequences. The study shows that the CTR of the AM-mode was 7.5 dB higher than the B-mode on average, while the PI-mode is 5.9 dB lower than the B-mode. This indicates that contrast-enhancement can be achieved with a CMUT when a suitable image sequence is chosen.
Language: | English |
---|---|
Publisher: | IEEE |
Year: | 2020 |
Pages: | 1-4 |
Proceedings: | 2020 IEEE International Ultrasonics Symposium |
ISBN: | 1728154480 , 1728154499 , 9781728154480 and 9781728154497 |
ISSN: | 19485727 |
Types: | Conference paper |
DOI: | 10.1109/IUS46767.2020.9251326 |
ORCIDs: | Øygard, Sigrid H. , Ommen, Martin L. , Engholm, Mathias , Diederichsen, Søren E. , Thomsen, Erik V. , Stuart, Matthias Bo and Jensen, Jørgen Arendt |