Conference paper
Adaptive Receive and Transmit Apodization for Synthetic Aperture Ultrasound Imaging
This paper suggests a framework for utilizing adaptive, data-dependent apodization weights on both the receiving and transmitting aperture for Synthetic Aperture (SA) ultrasound imaging. The suggested approach is based on the Minimum Variance (MV) beamformer and consists of two steps. A set of uniquely designed receive apodization weights are applied to pre-summed element data forming a set of adaptively weighted images; these are in SA literature conventionally referred to as low-resolution images.
The adaptive transmit apodization is obtained by applying MV across the full set of single emission images before summation. The method is investigated using simulated SA ultrasound data obtained using Field II. Data of 13 point targets distributed at depths from 40 mm to 70 mm, and a 5.5 MHz, 64-element linear array transducer have been used.
The investigation has shown that the introduction of adaptive apodization weights on the transmitting aperture provides a main-lobe reduction (estimated at -30 dB) by a factor of 1.8 compared to the method using adaptive apodization weights on the receiving aperture only.
Language: | English |
---|---|
Publisher: | IEEE |
Year: | 2009 |
Pages: | 1-4 |
Proceedings: | 2009 IEEE International Ultrasonics Symposium |
ISBN: | 142444389X , 142444389x , 1424443903 , 9781424443895 and 9781424443901 |
ISSN: | 19485719 , 19485727 and 10510117 |
Types: | Conference paper |
DOI: | 10.1109/ULTSYM.2009.5442035 |
ORCIDs: | Jensen, Jørgen Arendt |