Calibration of thin-film dosimeters irradiated with 80-120 kev electrons

From

Risø National Laboratory for Sustainable Energy, Technical University of Denmark1

A method for calibration of thin-film dosimeters irradiated with 80-120keV electrons has been developed. The method is based on measurement of dose with a totally absorbing graphite calorimeter, and conversion of dose in the graphite calorimeter to dose in the film dosimeter by Monte Carlo calculations.

A thermal model was developed to estimate the temperature contributions from the air above the calorimeter that is heated by the electron beam. As an example, Riso B3 thin-film dosimeters were calibrated by 80-120 keV electron irradiation and compared with a calibration carried out at 10 MeV electron irradiation.

The two calibrations were found to be equal within the estimated uncertainties of +/-10% at 1 s.d. (C) 2004 Elsevier Ltd. All rights reserved.

Language: English
Year: 2004
Pages: 355-359
Proceedings: 13th International Meeting on Radiation Processing
ISSN: 0969806x and 18790895
Types: Journal article and Conference paper
DOI: 10.1016/j.radphyschem.2004.03.032
ORCIDs: Helt-Hansen, J. and Miller, A.