Duplex synthetic aperture imaging with tissue motion compensation

From

Department of Electrical Engineering, Technical University of Denmark1

Center for Fast Ultrasound Imaging, Centers, Technical University of Denmark2

This paper investigates a method for tissue motion estimation and compensation in synthetic transmits aperture imaging. The approach finds the tissue velocity and the direction of the motion at very tissue region by cross-correlating high resolution lines beamformed along multiple directions at each image points.

Compensation is applied in the beamformer by tracking the image points using the velocity and angle estimates from the closest estimation point. Simulation results using Field II show nearly perfect motion compensation with no appreciable difference in contrast resolution after compensation. Phantom measurements show similar performance with differences in contrast resolution of 29% and 0.61% before and after compensation, respectively.

Language: English
Publisher: IEEE
Year: 2003
Pages: 1569-1573
Proceedings: 2003 IEEE Ultrasonics Symposium
Series: I E E E International Ultrasonics Symposium. Proceedings
ISBN: 0780379225 and 9780780379220
ISSN: 15513025 and 10510117
Types: Conference paper
DOI: 10.1109/ULTSYM.2003.1293207
ORCIDs: Jensen, Jørgen Arendt

DTU users get better search results including licensed content and discounts on order fees.

Log in as DTU user

Access

Analysis