A fast and reliable way to establish fluidic connections to planar microchips

From

ChemLabChip Group, LabChip Section, Department of Micro- and Nanotechnology, Technical University of Denmark1

LabChip Section, Department of Micro- and Nanotechnology, Technical University of Denmark2

Department of Micro- and Nanotechnology, Technical University of Denmark3

Polymeric Enabling Microsystems Group, Polymer Micro and Nano Engineering Section, Department of Micro- and Nanotechnology, Technical University of Denmark4

Polymer Micro and Nano Engineering Section, Department of Micro- and Nanotechnology, Technical University of Denmark5

In this work, we present a non-permanent method to connect microfluidic devices. The approach uses short flexible tubes that are plugged into bottom-flat holes and ensure fast and reliable interconnections. The small available dimensions allow the tube to be directly attached to the side of planar microchips.

A theoretical model to estimate the maximum applicable pressure was developed, and verified with experimental data. Furthermore, the tube connections were compared to other non-permanent interconnection types.

Language: English
Year: 2007
Pages: 98-103
ISSN: 13616439 and 09601317
Types: Journal article
DOI: 10.1088/0960-1317/17/1/013