About

Log in?

DTU users get better search results including licensed content and discounts on order fees.

Anyone can log in and get personalized features such as favorites, tags and feeds.

Log in as DTU user Log in as non-DTU user No thanks

DTU Findit

Journal article

A sensitive, specific and simple loop mediated isothermal amplification method for rapid detection of campylobacter spp. In broiler production

From

Department of Biotechnology and Biomedicine, Technical University of Denmark1

National Food Institute, Technical University of Denmark2

Research Group for Food Microbiology and Hygiene, National Food Institute, Technical University of Denmark3

Nanomaterials and Nanobiosensors, Experimental & Translational Immunology, Department of Health Technology, Technical University of Denmark4

Experimental & Translational Immunology, Department of Health Technology, Technical University of Denmark5

Department of Health Technology, Technical University of Denmark6

Campylobacteriosis is one of the most common foodborne diseases worldwide. Two Campylobacter species – C. jejuni and C. coli in poultry and poultry products are considered to be the main source of human campylobacteriosis. Therefore, studying Campylobacter status in poultry flocks is needed to prevent transmission of disease and reduce human risk, health cost, and economic losses.

In this study, we adapted and used a Loop-Mediated Isothermal Amplification (LAMP) assay for specific, sensitive, simple and cost-effective rapid detection of C. jejuni and C. coli in the poultry production chain. Amplified LAMP products were detected using a small, low-cost portable commercial blue LED transilluminator and a direct visual detection strategy was demonstrated.

By using optimized conditions for amplification a limit of detection (LOD) of 50 CFU/ml was achieved for testing of C. jejuni and C. coli in spiked chicken feces without enrichment. The method took 60–70 min from receiving the samples to the final results (including 30 min for amplification). The optimized LAMP showed a relative accuracy of 98.4%, a specificity of 97.9%, and a sensitivity of 100% in comparison to real-time PCR method.

Cohen’s kappa index also showed an excellent agreement (0.94) between the two methods. The results showed that the method is specific, sensitive and is suitable to develop for rapid detection of Campylobacter spp. at poultry production.

Language: English
Publisher: Frontiers Media S.A.
Year: 2019
Pages: 2443
ISSN: 1664302x
Types: Journal article
DOI: 10.3389/fmicb.2019.02443
ORCIDs: Than Linh, Quyen , Aaydha Chidambara, Vinayaka , Anh, Tien Ngo , Sun, Yi , Bang, Dang Duong and Wolff, Anders

DTU users get better search results including licensed content and discounts on order fees.

Log in as DTU user

Access

Analysis