Upgrade of an old Raman Spectrometer

From

Department of Chemistry, Technical University of Denmark1

Department of Chemical and Biochemical Engineering, Technical University of Denmark2

Improvement of a conventional Jeol Raman spectrometer with a single channel photo multiplier detector is described. New optical components (fibres, mirror, lens and CCD detector) have been chosen to design a high quality and easy-to-use instrument. Tests have shown that with this modified spectrometer Raman spectra can be acquired of a quality comparable to the spectra obtained previously, but the time needed to obtain a spectrum is markedly reduced.

Selected test spectra and a simple calibration procedure to obtain the wavenumber values from the band CCD pixel position are presented.