Accelerated equilibrium sampling of hydrophobic organic chemicals in solid matrices: A proof of concept on how to reach equilibrium for PCBs within 1 day

In Chemosphere 2019, Volume 237, pp. 124537
From

Technical University of Denmark1

Department of Environmental Engineering, Technical University of Denmark2

Environmental Fate & Effect of Chemicals, Department of Environmental Engineering, Technical University of Denmark3

University of Insubria4

Equilibrium sampling of hydrophobic organic chemicals (HOCs) is increasingly used to measure freely dissolved concentrations and chemical activities in sediments and soils. However, for the most hydrophobic chemicals (Log Kow > 6) such equilibrium sampling requires often very long sampling times in the order of weeks to months.

The aim of the present study was to explore two strategies for markedly increasing the HOC mass transfer from matrix to sampler with the overall goal to shorten equilibration times down to a few hours. Two Solid Phase Microextraction (SPME) approaches were thus developed and tested in sediment and soil contaminated by polychlorinated biphenyls (PCBs).

In the first method, the SPME fiber was immersed directly in the aqueous suspension of the sample under vigorous agitation. In the second method equilibration took place via the headspace and was accelerated by elevating the temperature. Headspace-SPME at 80 °C provided fast equilibration within approximately 2 h without contacting the sample and thus avoiding fiber fouling.

Both SPME methods were calibrated by passive dosing from preloaded silicone rods and yielded similar results, supporting the validity of HS-SPME at elevated temperatures on a proof of principle level. Finally, by using 13C labelled PCB standards, total concentrations were simultaneously measured, which in turn allowed calculation of matrix–water distribution coefficients.

Language: English
Year: 2019
Pages: 124537
ISSN: 18791298 and 00456535
Types: Journal article
DOI: 10.1016/j.chemosphere.2019.124537
ORCIDs: Knudsmark Sjøholm, Karina , 0000-0001-9284-2763 and Mayer, Philipp

DTU users get better search results including licensed content and discounts on order fees.

Log in as DTU user

Access

Analysis