Comparison of Continuous Glucose Monitoring Accuracy Between Abdominal and Upper Arm Insertion Sites

From

University of Copenhagen1

Scientific Computing, Department of Applied Mathematics and Computer Science, Technical University of Denmark2

Department of Applied Mathematics and Computer Science, Technical University of Denmark3

Background: The aim was to compare the accuracy of the Dexcom® G4 Platinum continuous glucose monitor (CGM) sensor inserted on the upper arm and the abdomen in adults. Methods: Fourteen adults with type 1 diabetes wore two CGMs, one placed on the upper arm and one placed on the abdomen. Three in-clinic visits of 5 h with YSI (2300 STAT, Yellow Springs Instrument) measurements as comparator were performed.

Each visit was followed by 4 days with seven-point self-monitoring of blood glucose (SMBG) in free-living conditions. Accuracy analyses on the paired CGM-YSI and CGM-SMBG measurements of the two CGM sensors were performed. Results: Using YSI as comparator, the overall Mean Absolute Relative Difference (MARD) for the CGMabd was 12.3% and CGMarm was 12.0%.

The percentage of the CGM measurements in zone A of Clarke error grid analysis for the CGMabd was 85.6% and CGMarm was 86.0%. The hypoglycemia sensitivity for the CGMabd and CGMarm was 69.3%. Using SMBG as comparator, the overall MARD for the CGMabd was 12.5% and CGMarm was 12.0%. The percentage of the CGM measurements in zone A for the CGMabd was 84.1% and the CGMarm was 85.0%.

The hypoglycemia sensitivity for the CGMabd was 60.0% and the CGMarm was 71.1%. All the P-values from the comparisons between the accuracy of CGMabd and CGMarm were >0.05. Conclusion: The accuracy of a Dexcom G4 Platinum CGM sensor placed on the upper arm was not different from the accuracy of the sensor placed on the abdomen in adults with type 1 diabetes.

Language: English
Publisher: Mary Ann Liebert, Inc., publishers
Year: 2019
Pages: 295-302
ISSN: 15578593 and 15209156
Types: Journal article
DOI: 10.1089/dia.2019.0014
ORCIDs: Mahmoudi, Zeinab and Jørgensen, John Bagterp

DTU users get better search results including licensed content and discounts on order fees.

Log in as DTU user

Access

Analysis