Journal article
A Portable Single Axis Magnetic Gradiometer
Department of Electrical Engineering, Technical University of Denmark1
Department of Automation, Technical University of Denmark2
Solar System Physics, National Space Institute, Technical University of Denmark3
National Space Institute, Technical University of Denmark4
Measurement and Instrumentation Systems, National Space Institute, Technical University of Denmark5
The single axis magnetic gradiometer based on two compact detector compensation (CDC) fluxgate ringcore sensors separated 20 cm is described. Despite its high stability and precision better than 1 nT, the calibration procedures are not straightforward. Firstly, the mono-axial measurement does not provide vector information about the magnetic field.
Secondly, one of the sensors measures the ambient magnetic field and is used to compensate for the main field at both sensors. Several methods have been developed for characterization of the 2 gradiometer, and the calibration of the gradient measurements is achieved by using a magnetic dipole of strength 2 mAm(2).
In a coil facility, the gradient can be determined with an accuracy of 0.3 nT/m(RMS).
Language: | English |
---|---|
Year: | 2001 |
Pages: | 185-196 |
ISSN: | 18733069 and 09244247 |
Types: | Journal article |
DOI: | 10.1016/S0924-4247(01)00656-2 |
ORCIDs: | Merayo, José M.G. , Petersen, Jan Raagaard and Brauer, Peter |