About

Log in?

DTU users get better search results including licensed content and discounts on order fees.

Anyone can log in and get personalized features such as favorites, tags and feeds.

Log in as DTU user Log in as non-DTU user No thanks

DTU Findit

Journal article

Aberration corrected and monochromated environmental transmission electron microscopy: challenges and prospects for materials science

From

Center for Electron Nanoscopy, Technical University of Denmark1

The latest generation of environmental transmission electron microscopes incorporates aberration correctors and monochromators, allowing studies of chemical reactions and growth processes with improved spatial resolution and spectral sensitivity. Here, we describe the performance of such an instrument using examples taken from studies of fuel cells and supported catalysts.

We discuss the challenges involved in performing in situ gas reaction experiments quantitatively and reliably and we highlight the care required to understand the effect of the electron beam on dynamic experiments performed in the electron microscope.

Language: English
Year: 2010
Pages: 1338-1344
ISSN: 17432847 and 02670836
Types: Journal article
DOI: 10.1179/026708310X12756557336355
ORCIDs: Hansen, Thomas Willum and Wagner, Jakob Birkedal

DTU users get better search results including licensed content and discounts on order fees.

Log in as DTU user

Access

Analysis