Journal article

Fast differential scanning calorimetry of liquid samples with chips

From

Xensor Integration B.V.1

Department of Micro- and Nanotechnology, Technical University of Denmark2

Based on a modified version of standard chips for fast differential scanning calorimetry, DSC of liquid samples has been performed at temperature scan rates of up to 1000 °C/s. This paper describes experimental results with the protein lysozyme, bovine serum, and olive oil. The heating and cooling rate of the sensor is measured for temperature scan rates of up to 1300 °C/s with water and 2-butanol, in the temperature range of −90 °C/s to +130 °C/s.

The lysozyme is measured at temperature scan rates varying from 10 °C/s to 400 °C/s and in concentrations between 0.1% and 10% protein by weight. The bovine serum measurements show two main peaks, in good agreement with standard DSC measurements. Olive oil has been measured, with good agreement for the cooling curve and qualitative agreement for the heater curve, compared to DSC measurements.

Language: English
Year: 2015
Pages: 162-171
ISSN: 1872762x and 00406031
Types: Journal article
DOI: 10.1016/j.tca.2014.07.013