About

Log in?

DTU users get better search results including licensed content and discounts on order fees.

Anyone can log in and get personalized features such as favorites, tags and feeds.

Log in as DTU user Log in as non-DTU user No thanks

DTU Findit

Journal article

Evaluation of pharmacokinetic model designs for subcutaneous infusion of insulin aspart

From

University of Canterbury1

University of Copenhagen2

Department of Applied Mathematics and Computer Science, Technical University of Denmark3

Scientific Computing, Department of Applied Mathematics and Computer Science, Technical University of Denmark4

Dynamical Systems, Department of Applied Mathematics and Computer Science, Technical University of Denmark5

Effective mathematical modelling of continuous subcutaneous infusion pharmacokinetics should aid understanding and control in insulin therapy. Thorough analysis of candidate model performance is important for selecting the appropriate models. Eight candidate models for insulin pharmacokinetics included a range of modelled behaviours, parameters and complexity.

The models were compared using clinical data from subjects with type 1 diabetes with continuous subcutaneous insulin infusion. Performance of the models was compared through several analyses: R2 for goodness of fit; the Akaike Information Criterion; a bootstrap analysis for practical identifiability; a simulation exercise for predictability.

The simplest model fit poorly to the data (R2 = 0.53), had the highest Akaike score, and worst prediction. Goodness of fit improved with increasing model complexity (R2 = 0.85–0.92) but Akaike scores were similar for these models. Complexity increased practical non-identifiability, where small changes in the dataset caused large variation (CV > 10%) in identified parameters in the most complex models.

Best prediction was achieved in a relatively simple model. Some model complexity was necessary to achieve good data fit but further complexity introduced practical non-identifiability and worsened prediction capability. The best model used two linear subcutaneous compartments, an interstitial and plasma compartment, and two identified variables for interstitial clearance and subcutaneous transfer rate.

This model had optimal performance trade-off with reasonable fit (R2 = 0.85) and parameterisation, and best prediction and practical identifiability (CV < 2%).

Language: English
Publisher: Springer US
Year: 2017
Pages: 477-489
ISSN: 15738744 and 1567567x
Types: Journal article
DOI: 10.1007/s10928-017-9535-z
ORCIDs: 0000-0003-1661-2573 , 0000-0003-1620-8271 , Jørgensen, John Bagterp and Madsen, Henrik

DTU users get better search results including licensed content and discounts on order fees.

Log in as DTU user

Access

Analysis