About

Log in?

DTU users get better search results including licensed content and discounts on order fees.

Anyone can log in and get personalized features such as favorites, tags and feeds.

Log in as DTU user Log in as non-DTU user No thanks

DTU Findit

Journal article

Adaptive control in an artificial pancreas for people with type 1 diabetes

From

Department of Applied Mathematics and Computer Science, Technical University of Denmark1

Scientific Computing, Department of Applied Mathematics and Computer Science, Technical University of Denmark2

Copenhagen University Hospital Herlev and Gentofte3

Dynamical Systems, Department of Applied Mathematics and Computer Science, Technical University of Denmark4

Copenhagen Center for Health Technology, Centers, Technical University of Denmark5

Center for Energy Resources Engineering, Centers, Technical University of Denmark6

In this paper, we discuss overnight blood glucose stabilization in patients with type 1 diabetes using a Model Predictive Controller (MPC). We compute the model parameters in the MPC using a simple and systematic method based on a priori available patient information. We describe and compare 3 different model structures.

The first model structure is an autoregressive integrated moving average with exogenous input (ARIMAX) structure. The second model structure is an autoregressive moving average with exogenous input (ARMAX) model, i.e. a model without an integrator. The third model structure is an adaptive ARMAX model in which we use a recursive extended least squares (RELS) method to estimate parameters of the stochastic part.

In addition, we describe some safety layers in the control algorithm that improve the controller robustness and reduce the risk of hypoglycemia. We test and compare our control strategies using a virtual clinic of 100 randomly generated patients with a representative inter-subject variability. This virtual clinic is based on the Hovorka model.

We consider the case where only half of the meal bolus is administered at mealtime, and the case where the insulin sensitivity increases during the night. The numerical results suggest that the use of an integrator leads to higher occurrence of hypoglycemia than for the controllers without the integrator.

Compared to other control strategies, the adaptive MPC reduces both the time spent in hypoglycemia and the time spent in hyperglycemia.

Language: English
Year: 2017
Pages: 332-342
ISSN: 18736939 and 09670661
Types: Journal article
DOI: 10.1016/j.conengprac.2016.01.003
ORCIDs: Boiroux, Dimitri , Poulsen, Niels Kjølstad , Madsen, Henrik and Jørgensen, John Bagterp

DTU users get better search results including licensed content and discounts on order fees.

Log in as DTU user

Access

Analysis