About

Log in?

DTU users get better search results including licensed content and discounts on order fees.

Anyone can log in and get personalized features such as favorites, tags and feeds.

Log in as DTU user Log in as non-DTU user No thanks

DTU Findit

Journal article

Ultrasensitive Real-Time Rolling Circle Amplification Detection Enhanced by Nicking-Induced Tandem-Acting Polymerases

From

Department of Health Technology, Technical University of Denmark1

Nano and Bio-physical Systems, Department of Health Technology, Technical University of Denmark2

Magnetic Systems, Nano and Bio-physical Systems, Department of Health Technology, Technical University of Denmark3

BluSense Diagnostics IVS4

Padlock probe ligation-based rolling circle amplification (RCA) can distinguish single-nucleotide variants, which is promising for the detection of drug-resistance mutations in, e.g., Mycobacterium tuberculosis (Mtb). However, the clinical application of conventional linear RCA is restricted by its unsatisfactory picomolar-level limit of detection (LOD).

Herein, we demonstrate the mechanism of a nicking-enhanced RCA (NickRCA) strategy that allows several polymerases to act simultaneously on the same looped template, generating single-stranded amplicon monomers. Limiting factors of NickRCA are investigated and controlled for higher amplification efficiency.

Thereafter, we describe a NickRCA-based magnetic nanoparticle (MNP) dimer formation strategy combined with a real-time optomagnetic sensor monitoring MNP dimers. The proposed methodology is applied for the detection of a common Mtb rifampicin-resistance mutation, rpoB 531 (TCG/TTG). Without additional operation steps, an LOD of 15 fM target DNA is achieved with a total assay time of ca. 100 min.

Moreover, the proposed biosensor holds the advantages of single-nucleotide mutation discrimination and the robustness to quantify targets in 10% serum samples. NickRCA produces short single-stranded monomers instead of the DNA coils produced in conventional RCA, which makes it more convenient for downstream operation, immobilization or detection, thus being applicable with different molecular tools and biosensors.

Language: English
Publisher: American Chemical Society
Year: 2019
Pages: 10102-10109
ISSN: 15206882 , 00032700 and 15204782
Types: Journal article
DOI: 10.1021/acs.analchem.9b02073
ORCIDs: Tian, Bo , 0000-0002-7515-4026 and Hansen, Mikkel Fougt

DTU users get better search results including licensed content and discounts on order fees.

Log in as DTU user

Access

Analysis