Free-field correction values for Interacoustics DD 45 supra-aural audiometric earphones

From

Hearing Systems, Department of Electrical Engineering, Technical University of Denmark1

Department of Electrical Engineering, Technical University of Denmark2

This paper report free-field correction values for the Interacoustics DD 45 audiometric earphones. The free-field correction values for earphones provide the loudness based equivalence to loudspeaker presentation. Correction values are especially used for the calibration of audiometric equipment for speech audiometry performed with headphones.

Calibration values may be found in e.g. the ISO 389 series of standards. The free-field correction values were determined by means of loudness balance measurements of one-third octave noises (centre frequencies 125 Hz to 8000 Hz) presented alternately from a loudspeaker in a free field and from the earphones.

The procedure was essentially in accordance with the free-field frequency response procedure described in IEC 60268-7: Headphones and earphones. The study sample consisted of four earphones and 14 test subjects. Free field correction values are reported for the acoustic coupler IEC 60318-3 (NBS 9-A) and for the ear simulator IEC 60318-1.

The results are in good agreement with the results of another independent investigation. The reported free-field correction values may be used as part of the basis for future standardization of the DD 45 earphone.