Conference paper
A microfluidic control system with re-usable micropump/valve actuator and injection moulded disposable polymer lab-on-a-slide
A microfluidic control system consisting of micropump/valves with a re-usable pneumatic actuator and a disposable polymer lab-on-a-slide is presented. The lab-on-a-slide was fabricated using low cost methods, such as injection moulding of TOPAS® cyclic olefin copolymer (COC) slide, lamination of different layers of polymer, and ultrasonic welding of TOPAS® lid to the slide.
The re-usable pneumatic actuator not only simplifies the design of the lab-on-a-slide and reduces the fabrication cost, but also reduces the possibility of cross contamination during replacement of the disposable lab-on-a-slide. The experiment results show the micropump is capable of providing a wide range of flow rate (20–300 µL/min), and microvalves provides a reliable fluidic control in the different parts of the system.
Language: | English |
---|---|
Publisher: | IEEE |
Year: | 2011 |
Pages: | 1244-1247 |
Proceedings: | 16th International Solid-State Sensors, Actuators and Microsystems Conference |
ISBN: | 145770157X , 145770157x , 9781457701573 , 1457701553 , 1457701561 , 9781457701559 and 9781457701566 |
Types: | Conference paper |
DOI: | 10.1109/TRANSDUCERS.2011.5969452 |
ORCIDs: | Bu, Minqiang , Yi, Sun and Wolff, Anders |
COC slide Disposable lab-on-a-slide Micropump and microvalve Re-usable actuator TOPAS cyclic olefin copolymer slide cross contamination flow control injection moulded disposable polymer lab-on-a-slide injection moulding microfluidic control system microfluidics micropumps microvalves pneumatic actuators polymer blends reusable micropump-valve actuator reusable pneumatic actuator ultrasonic welding