Conference paper

Microstrip Resonator for High Field MRI with Capacitor-Segmented Strip and Ground Plane

In Proceedings of the International Society for Magnetic Resonance in Medicine — 2017
From

Department of Electrical Engineering, Technical University of Denmark1

Center for Magnetic Resonance, Department of Electrical Engineering, Technical University of Denmark2

Electromagnetic Systems, Department of Electrical Engineering, Technical University of Denmark3

Center for Hyperpolarization in Magnetic Resonance, Centers, Technical University of Denmark4

Copenhagen University Hospital Herlev and Gentofte5

High field MRI coils are often based on transmission line resonators. Due to relatively short wavelength of RF fields, such coils produce uneven field patterns. Here we show, that it is possible to manipulate magnetic field patterns of microstrip resonators in both planes (sagittal and transverse) segmenting stripe and ground plane of the resonator with series capacitors.

The design equations for capacitors providing symmetric current distribution are derived. The performance of two types of segmented resonators are investigated experimentally. To authors’ knowledge, a microstrip resonator, where both, strip and ground plane are capacitor-segmented, is shown here for the first time.

Language: English
Publisher: The International Society for Magnetic Resonance in Medicine
Year: 2017
Proceedings: ISMRM 25th Annual Meeting & Exhibition
Types: Conference paper
ORCIDs: Zhurbenko, Vitaliy and Petersen, Esben Thade

DTU users get better search results including licensed content and discounts on order fees.

Log in as DTU user

Access

Analysis