Conference paper
Protocol for multi-site quantitative evaluation of 13C radio frequency coils
GE Healthcare1
University of Texas Southwestern Medical Center2
Magnetic Resonance, Department of Health Technology, Technical University of Denmark3
Department of Health Technology, Technical University of Denmark4
Center for Hyperpolarization in Magnetic Resonance, Centers, Technical University of Denmark5
We present a protocol for measurement of SNR profile of 13C RF coils for clinical imaging systems. This protocol makes use of standard, vendor-provided pulse sequences as well as the natural abundance 13CH3 resonance of the dimethyl silicone (DMS) phantoms which are widely distributed. We also provide an open source code for processing and analysis.
Language: | English |
---|---|
Year: | 2019 |
Proceedings: | ISMRM 27th Annual Meeting & Exhibition |
Types: | Conference paper |
ORCIDs: | Hansen, Rie Beck |